• Products
  • Support
  • Technical information
  • Download
  • Corporate information

Home > Products > Data Acquisition Systems > Instrumentation Amplifiers > WGA-900A

WGA-900A NEW

Instrumentation Amplifier
  • Recommended sensors:
  • ロードセル
  • 圧力変換器
  • 変位変換器
WGA-900A

Enables Checking Waveform through Easy Operation

Features

  • Easy to see display numbers of 5 digits (±99999)
  • Wide measuring range up to ±3.2mV/V
  • Comparator in 5 points enables delicate monitoring and controlling
  • Display of waveform
  • Easy calibration even for TEDS compatible senors
  • Easy operation using touch panels on with easy explanations. No need to see a operation manual.
  • SD card available as standard
  • High resolution and high speed sampling
  • Result of comparison will be showed on different colors
  • Analog-hold circuit is used for detecting peak/bottom value
  • MONITOR – output, RS-232C and I/O port is equipped as standard.
  • Optional BCD-output, D/A output or RS-485 card are available.

PDF download

Specifications

Standard specification

WGA-900A-0
Number of Measuring Channels 1
Applicable Transducer Strain gage transducer
Applicable Bridge Resistance 87.5 Ω to 1 KΩ (Up to four 350 Ω transducers can be connected in parallel.)
TEDS Compatible Interface: Compatible to IEEE1451.4 Mixed ModeTransducer Interface Class 2.Applicable transducer: Should have the informationaccording to IEEE template No. 33.(Cable length: 30 m or less)
Excitation Voltage 10 or 2V DC
Input Range ±3.2 mV/V (including zero adjustment range)
Nonlinearity Within (±0.02%FS +1 digit)
Stability Zero : Within ±0.25μVRTI/°C
Sensitivity : Within ±0.005%/°C
Peak/Bottom Detection Detecting method : Analog circuit and digital holdin combination/Digital hold onlyFrequency response : DC to 1 kHz (+1 dB, -2dB)
A/D Converter Sampling speed : 4000 times/secResolution: 24 bits
Analog Monitor Voltage output: ±(5 V ±200 mV)(load resistance 5 KΩ or more)
Indicators 3.5-inch STN color LCD,display area : 73.0 × 55.2 mm320 × 240 dots, touch panel
Indication ±99999Speed: 3 times/sec
Calibration Manual calibration : No-load zero calibration, sensitivity registeringcalibration, actual load calibration, unit
TEDS-based automatic calibration Partial calibration based on TEDS calibration data: TEDS calibration item
TEDS operation setting : TEDS reading operation, zero at TEDS  calibration time, TEDS information display
Smoothing Function Analog filter : 1, 30, 300 Hz and None(1 kHz or more)Roll-off characteristic: -12 dB/oct.
Minimum scale : 1, 2, 5, 10, 20, 50, 100
Moving average : None, 2, 4, 8, 16, 32, 64, 128, 256, 512, 1024,and 2048 times
Zero Compensation Zero at every preset compensation time(Conducts digital zero when the measuring valueis within the compensation range at every presetcompensation time.)Determination Time : 0.00 to 9.99 secCompensation Range : 0 to 99999
Least significant digit fixed at zero(Automatically changes the least significantdigit number to zero.)Setting Range: 0 to 9
Additional Value Setting range: ±99999
Sensor Output Value Transducer output (±3.2 mV/V) is measuredat 0.5-second intervals.
Measurement condition numbers 32 points (16 for control input) of measurement condition file can be saved. Capable of switching by the key operation, control input, and communication command.
Comparator Setting The number of points : 5
Type: extra high (HH), high (HI), OK, low (LO), extra low (LL)
Compared value : ±99999
Hysteresis Width : 0 to 9999
Using comparator can be set.
Comparison speed : 4000 times per second (Normal Comp. Mode)
Measurement Mode Operation mode : Normal, peak hold, block-specified peak hold,time-specified peak hold, bottom hold,block-specified bottom hold,time-specified bottom hold, arbitrary point hold,block peak-bottom, time peak-bottom,block arerage, time average
Detect times : Delay time, comparison mode and display modecan be set.
Waveform Display X-axis setting End point : 0.5, 1.0, 2.0, 5.0, 10.0 seconds
Y-axis setting Start point : -99999 to 99999End point : 250, 500, 1000, 2000, 5000, 10000,20000, 50000, 100000, 200000
Start mode of waveform, passed level, passed way, hold time of waveform Displays the waveform of the input variation regardless to the “Measure Mode Set” setting.
System Setting Key lock, setting value initialize,backlight illumination time, contrast, clock
Self-check Memory, channel
Operation Check Display, touch panel, control input/output,communication, BCD output, D/A output, SD card
Control Input Number of points : 9
Type : Zero command, hold command, reset command,waveform command, TEDS command,measurement condition select 0 to 3
Signal format : Non-voltage contact signal or open collector signal(12 VDC voltage and 5 mA current can be applied.)
Control Output Number of points: 10
Type : HH, HI, OK, LO, LL, healthy, abnormal channel,abnormal memory, communication error, SD
Output format : Open collector
Load capacity : 30 VDC, 20 mA
Communication Signaling system : RS-232C full duplex system
Transmission system : Asynchronous
Communication speed : 2400, 4800, 9600, and 19200 bps
Bit configuration
Data bit 7
Stop bit 1
Parity Odd number
Flow control Not compatible
SD Card Saving set value : Saves the all setting value (excluding thecalibration value) to the SD card.
Reading setting value : Reads the all setting value (excluding thecalibration value) from the SD card and rewritethose of the WGA-900A to the read one.
Waveform data editing : Browsing the waveform data,changing the name of the data,
Waveform data editing : deleting the data are available.
Format : Erase all data that are saved in the SD card (Quick format)available
Update : Capable of updating the program version that is savedin the SD card.
SD card type : Up to 2 GB; Non-SDHC-compliant
Power Supply 85 to 264 VAC, 50/60 Hz, 20 VA or less
Dimensions 100 mm (W) × 96 mm (H) × 135 mm (D)
Weight : Approx. 1.0 kg (excluding options)
Operating Temperature 0 to 40 °C, 20 to 85 %RH (noncondensing
Pannel cut Dimentions 92×92mm

Optional Accessories

AC power supply cable(P-23), SD card(2GB)

OPTION Specifications

with BCD Output

WGA-900A-1
Output Format : Open-collector
BCD data : 20 bits (4-bit×5-digit) EOC (End Of Conversion) : 1 bit
Format : Open-collector  Max load capacity : 30V DC, 20 mA (load resistance)
Input Number of points : 2
Type : Open collector or Non-voltage contact signalVoltage 12V DC and current 10 mA can be applied.
Porality logic,
EOC logic,
Data logic
Negative or positive logic
Transmission speed Approx. 16, 32, 64, and125 times/sec.

with D/AOutput

WGA-900A-2
OutputVoltage ±10 V (load resistance 2 kohm or more),arbitrary scaling is available.
Output Current 4 to 20 mA (load resistance 500 ohm or less)4 to 20 mA output is fixed when the voltage0 to 10 V is applied.
Insulation Voltage 250V AC for 1 minute (output voltage and outputcurrent are non-insulated)
Conversion Speed 2000 times/sec
Nonlinearity ±0.1%FS

with RS-485

WGA-900A-3
Signaling System RS-485 Half duplex system
RS-485 Half duplex system 2400, 4800, 9600, 19200 bps
Device ID Setting range : 1 to 99
Bit Structure Data bit : 7
Stop bit : 1
Parity : Odd number
Flow Control Not compatible

with BCD and D/A Output

WGA-900A-12
BCD/ Binary output
Output Format : Open-collector
BCD data : 20 bits (4-bit×5-digit) 1 bit Over : 1 bitEOC (End Of Conversion) : 1 bit
Format : Open-collectorMax load capacity : 30V DC, 20 mA (load resistance)
Input Number of points : 2
Type : Open collector or Non-voltage contactsignal Voltage 12V DC and current 10 mAcan be applied.
Transmission speed Approx. 16, 32, 64,and 125 times/sec.
Output Voltage ±10 V (load resistance 2 kohm or more),
arbitrary scaling is available.
Output Current 4 to 20 mA (load resistance 500 ohm or less)
4 to 20 mA output is fixed when the voltage0 to 10 V is applied.
Insulation Voltage 250V AC for I minute (output voltage and outputcurrent are non-insulated)
Conversion Speed 2000 times/sec
Nonlinearity ±0.1%FS

Back to Top

Registration

Please register for our website support when downloading files or documents.

New registration

Inquiries from the Web