PCD-330B-F,PCD-331B-F

- Recommended sensors:








4 channels per unit, up to 4 units can be connected for 16 channel measurement. Enables simultaneous measurement of strain and voltage signals on PC.
Connected to the PC via the USB port, the PCD-300 series makes the PC a measuring instrument by connecting via USB port. Chose the PCD-300B interface for strain gages to measure strain/stress, force, pressure, acceleration, and displacement. Chose the PCD-320A with voltage-output interface to measure various signals detected in voltage. Connect strain gages or voltage output sensors to input terminals and begin dynamic data acquisition software DCS. Each sensor interface provides 4 channels and can configure up to 16 channels by connecting 4 units.
Features
●There are five models of input adaptors.
● Dynamic data acquisition software DCS-100A comes standard with PCD-330B-F.
● 24-bit A/D converter
● Low-pass filter mounted , DCS-100A provided standard provision.
● TEDS compatible
※For data acquisition software (DCS-100A),
refer to here
.
※For TEDS, refer to technical Notes
PDF download
- PCD-300 (869.23 kB)
Specifications
| Number of channel | 4/台(最大4台16チャネルまで同期運転可能) |
| Synchronous measurement | Up to 4 units by 16 channels |
| Input adapter |
UI-10A (Strain gage transducer with NDIS connector) (When the voltage is measured, FV-1A (Option) is installed.) |
| UI -11A (Strain gage, Clamp-style terminal block) | |
| UI-15A (Strain gage, Clamp-style terminal block with operating levers) | |
| UI-16A (Strain gage, Clamp-style terminal block with quick-fitting locks) | |
| UI-30A (Voltage signal, BNC) | |
| A-D Conversion | Max. 24bit |
| Sampling Frequency | Up to 10kHz (Simultaneous sampling of 4 channels) |
| TEDS Compatibility | With the UI-10A/UI-11A input adapter mounted |
| Holding Setting Conditions | Range, Balance Adjustment are stored in internal nonvolatile memory. |
| Interface | USB1.1 |
| Operating Temperature/Humidity Range | 0 to 40℃, 20 to 80% RH (noncondensing) |
| Vibration Resistance | ±29.42m/s2: (3G), 5 to 200Hz |
| (12 cycle/ axis, 10 min/cycle) | |
| Power Supply | DC 11 to 30V |
| 100 to 240 VAC (Accessory UI318-12 AC adapter should be used.) | |
| Current Consumption | 0.9 A or less (12 VDC) |
| Dimensions | 265.2(W) x 26.7(H) x 215(D) mm (excluding protrusions) |
| Approx. 1.1 kg (excluding accessory AC adapter;with UI-10A mounted) | |
| EMC | Conforms to IEC61326-1, Class A Equipment |
Standard Accessories
· UI318-12 AC adapter
· N-38 USB cable 1 m long
· P-72 grounding conductor 5 m long
· CD-R (contains DCS-100A dynamic data acquisition software,Instruction Manual, etc.)
Optional Accessories
· P-73 DC power cable 2 m long (11 to 30 VDC)
· Connection cables N-97 10 cm long, U-115 50 cm long,U-116 1 m long, U-117 2 m long,U-118 5 m long
· N-90 synchronous cable 25 cm long (required for unit-to-unit connection)
· N-39 USB cable 2m long
· SA-34A AC adapter for 4 units
· Input adapters (the mainframe should have one of them mounted.)UI-10A, UI-11A, UI-15A, UI-16A, UI-30,(FV-1A Voltage conversion adapter)
|
UI-10A |
UI-11A |
UI-15A |
UI-16A |
UI-30A |
|
FV-1A |
Strain/Stress measurement Mode
| Applicable Sensors | Strain gages, Strain gage transducers |
| Applicable Gage Resistance | 120Ω(1-gage quarter bridge and 2-gage half bridge) |
| 120Ω to 1 kΩ (4-gage full bridge) | |
| Input Mode | Balanced differential |
| Bridge Excitation | 2 VAC rms, carrier 1 kHz sine wave |
| Balance Adjustment Range | Resistance ±2% (±10000μm/m) or |
| more Capacitance 5000 pF or more | |
| Balance Adjustment Method | Resistance True electron auto balance |
| method Capacitance CST | |
| (capacitance self tracking) method | |
| Nonlinearity | Within ±0.1% FS |
| Gage Factor | 2.00 fixed |
| Range Selection | 8 steps of 200, 500, 1000, 2000, 5000, 10000,20000μm/m and OFF |
| Accuracy within ±0.5% FS | |
| Frequency Response Range | DC to 200 Hz, deviation ±10% |
| Low-pass Filter | Second order Butterworth |
| Cutoff frequency : 4 steps of 10, 30, 100 Hz and Flat | |
| Amplitude ratio at cutoff point : -(3±1) dB | |
| Attenuation : -(12±1) dB/oct. | |
| Zero Stability | Within ±1μm/m8 h, within ±0.2μm/m℃ |
| Sensitivity Stability | Within 0.3%/8 h, within 0.05%/℃ |
| Withstand voltage | 250VAC for 1 minute between input and chassis |
Voltage measurement Mode
| Applicable Sensors/Instruments | Voltage-output transducers,Voltage-output conditioners |
| Input Mode | Unbalanced |
| Range | 7 steps of 1, 2, 5, 10, 20, 50 V and OFF |
| Accuracy | Within 0.2% FS |
| Frequency Response Range |
DC to 1 kHz Deviation : Within -1/±3 dB |
| High-pass filter | 2 steps of 0.2Hz and Flat |
| Low-pass Filter | Second order Butterworth |
| Cutoff frequency : 5 steps of 10, 30, 100, 300 Hz and Flat | |
| Amplitude ratio at cutoff point : -(3±1) dB | |
| Attenuation: -(12±1) dB/oct. | |
| Zero Stability | Within ±0.1%FS/8h, within ±0.01%FS/℃ |
| Sensitivity Stability | Within ±0.3% /8h, within ±0.03% /℃ |