UCAM-60B,UCAM-65B

Data Loggers
  • Recommended sensors:
  • ひずみゲージ
  • カプセルゲージ
  • ロードセル
  • 圧力変換器
  • 変位変換器
  • 電圧
  • 熱電対
  • 土木用変換器
UCAM-60B
UCAM-65B

Can measure up to 20000 μm/m with a resolution as high as 0.1 μm/m. (※When using for Full bridge system)
The data logger UCAM-60B is an all-in-one measuring instrument developed in full pursuit of easier field measurement. Easy to operate keys, a bright readable display providing understandable presentation and a printer for immediate confirmation of measurement results. All these and more are incorporated in this compact unit to satisfy every need in field measurement. The UCAM-65B is a compact online data logger fully controlled through the PC. Both models can connect to, and simultaneously input signals from, strain gages, strain gage transducers, civil engineering transducers with temperature measuring function, potentiometer sensors, thermocouples and DC voltage-output instruments. They are also compatible with TEDS-installed sensors having information conforming to IEEE template No. 33. While measurement in a maximum 30 channels is possible with the mainframe only, external scanners enable measurement in a maximum 1000 channels. Measured results are stored in internal memory. And for easy data transfer to PC, measured results can also be saved in a flash ATA card or CF card inserted into the PC card slot. Furthermore, Ethernet LAN and RS-232C interfaces are provided standard for connection to the PC, and the control software UCS-60B enables the PC to not only control the UCAM-60B/65B but also perform data processing for rosette analysis, etc. in the field by directly collecting data.

Features

【UCAM-60B】

  • Easy to understand English presentation
  • Fluorescent display tube ensuring easy viewing in the field
  • Built-in thermal printer for immediate confirmation of measured results

【UCAM-65B】

  • Setting measuring conditions from PC and saving measured results to PC
  • Interval measurement possible with no PC connected

【Common to UCAM-60B and UCAM-65B】

  • Measurement up to 20,000 m/m with a resolution of 0.1 m/m (※When using for Full bridge system)
  • Scanning at 50 ms/channel with internal scanners
  • High-speed scanning at 20 ms/channel with internal scanners
  • Up to 30 channel measurement with internal scanners
  • Up to 1000 channel measurement with external scanners
  • PC card slot ensuring easy data collection
  • DC operated version for operation where no AC outlet is avalaible
  • Can automatically set the gage mode for each channel by detecting the channel mode corresponding to the connected strain gage or strain gage transducer.
  • TEDS compatible (with internal scanner USS-61B/62B/ 63B mounted; for TEDS, refer to page 8-21)
  • Control software UCS-60B (optional for UCAM-60B) enables control through PC connected via Ethernet LAN or RS-232C. (When connecting via Ethernet LAN, use a straight cable and LAN hub.)

(Note 1) For TEDS, refer to P.9-15.
(Note 2) When connecting via Ethernet LAN use a straight cable and LAN hub

PDF download

System Components

Data Logger(Main Unit)

Model Power Supply Control Software UCS-60B
UCAM-60B-AC AC only Optional
UCAM-60B-DC DC only Optional
UCAM-65B-AC AC only Standard
UCAM-65B-AC-0 Optional
UCAM-65B-DC DC only Standard
UCAM-65B-DC-0 Optional
Dedicated Scanners USS-61B for general purpose
USS-62B for general purpose with NDIS connectors*1
USS-63B for civil engineering with lightning arrester
The main unit can accommodate up to 3 dedicated scanners.
External Scanners The main unit can connect to the following scanners via the optional scanner interface.
USB-70 series via scanner interface USI-67A
Scanner Interfaces USI-67A for USB-70 series
External Input/Output Unit UIO-60A
Control Software UCS-60B

*1. TEDS-compatible function is made effective by connecting

TEDS-installed sensor through NDIS connector.

Specifications

Data Loggers

UCAM-60B、65B
Applicable Sensors Strain gages, strain gage transducers, civil engineering transducers with temperature measuring function, DC voltage-output or DC current-output instruments, potentiometer sensors,thermal sensors (thermocouples and platinum resistance thermometer bulbs)

 

Scanners and
Applicable Sensors

Dedicated

Scanner

Dedicated Scanner
General purpose Civilengineering
USB-70B-10/20 USB-70B-30
strain gages and strain gage transducers Quarter bridge method 120Ω
240Ω
350Ω

 

Quarter bridge method

(true-dummy method)

 

120Ω
240Ω

 

Half bridge method

60 to 1000

 

Active-dummy method
Active-active method
Common dummy method  

 

Full bridge method

60 to 1000 *3

 

Opposite side active method
Full-bridge method

Civil engineering 

transducers

Full bridge method 120 Constant-current excitation    

 

Full bridge method

350

 

Constant-current excitation
With temp. measuring function  
Voltage

DC voltage-

output

instruments

Current

DC current-

output

instruments

Temperature Thermocouples K(CA)
T(CC)
E(CRC)
J(IC)
R

 

Platinum

resistance

thermometer

bulbs

 

Pt100 (new JIS)  
JPt100 (old JIS)  
Potentiometer sensors
Built-in lightning arrester ●(*1)  
Scanner interface *3 Not required -67A  

 

*1. With USS-63B mounted.

*2. Use either of the scanner interfaces for

*3. 120 to 1000Ω in high-resolution mode.

 

Number of Measuring Channels Max. 30 with dedicated scanners
Max. 1000 with external scanners connected
Scanning Speed 50 ms/channel (standard mode)
280 ms/channel (high-resolution mode), individually selectable for
desired channels 20 ms/channel (high-speed mode),
collectively selectable for all channels of dedicated scanners
Line Frequency 50 Hz Zone 60 Hz Zone
Scanner
Dedicated scanner (standard mode) 50 ms/channel
Dedicated scanner (high-resolution mode) 280 ms/channel
Dedicated scanner (high-speed mode) 20 ms/channel
USB-70 series (standard mode only) 60 ms/channel 58.4 ms/channel

Note : Scanning speeds stated above are standard maximum speeds in respective modes. Besides these, the following speeds can be set for each individual channel: 0.28 s, 0.5 s, 1 s, 2 s, 5 s and 10 s

Scanning Speed

 

Standard Mode

(50 ms/channel)

 

 

High-Resolution Mode

(280 ms/channel)

 

 

High-Speed Mode

(20 ms/channel)

 

Applicable Sensor
Strain (gage & transducer)
Voltage/current-output sensor ×
Civil engineering transducer × ×
Temperature sensor (TC, Pt) × ×
Potentiometer sensor ×

 

Notes : 1. High-resolution mode and high-speed mode are selectable for dedicated anners only.

2. High-resolution or high-speed mode is available only with

4-gage method.

 

Operating Modes Real-time, 
Monitor, and Automatic
Measurement Functions Initial: Initial values are measured and stored in internal memory (except for temperatures measured by civil engineering transducers with temperature measuring function).
Original:Raw values are measured without subtraction of initial values.
Measure : Initial values are subtracted from original values (except for temperatures measured by civil engineering transducers with temperature measuring function).
Easy Measure : Auto zero balancing function is activated.
Note: The selected function is applied to all channels. Coefficient
Calculation Function Multiplication by calibration coefficients,calibration by TEDS, conversion of measured values to physical quantities,
scaling and correction.
Engineering Units 59 units

 

Automatic Measurement Functions
Trigger Measurement

A relative value (certain changing quantity) or an absolute value triggers measurement. In addition to the usual trigger function,

a variable trigger function is provided with which the trigger value changes at each step during measurement. With this special function,

a trigger value and the number of measurement times (repeat times) under the trigger condition can be registered for each step to perform

a series of automatic measurements in the order of steps.

The maximum number of steps available for setting is 15 and the number of repeat times may be a value selected from a range of 1 to 9999 or infinite.

Trigger channel : 1 desired channel
Trigger value : A desired real number of 6 effective figures or less Reference
Reference value : Amount of level shift to determine the first trigger value
(selected from the same range as for the trigger values)
Number of repeat times : 1 to 9999 (0 for infinite times)
Number of measurement steps : Maximum 15
INTERVAL MEASUREMENT Measurement is automatically
performed at preset time intervals.
Starting time setting : Year/month/day and hour:minute:second
Interval setting : Days and hours:minutes:seconds (in a range
of 00 and 00:00:01 to 99 and 12:59:59)
Number of repeat times : 1 to 9999 (0 for infinite times)
Maximum number of steps : 15
TRIGGER INTERVAL MEASUREMENT Combination of trigger measurement and interval measurement.
Trigger value: Absolute value Number of measurement times: Max. 9999
Interval time: Available in a range of 1 second to 99 days 23 hours:59 minutes:59 seconds
Storage Function Internal memory,approx. 7 MB
Flash ATA card (optional); the capacity depends on the card.
Strain Measurement(Standard Mode)
Bridge excitation Constant voltage excitation : Approx. 2 or 5 VDC
Constant current excitation :
Approx. 5.7 mA (bridge resistance 350Ω)
(up to 5 km with a 4-conductor (0.5 mm 2 ) shielded cable)
Approx. 16.7 mA (bridge resistance 120Ω)
(up to 2 km with a 4-conductor (0.5 mm 2 ) shielded cable)
Scanning speed 50 ms/channel
Gage factor 2.00 fixed (coefficient calculation function enables 2.00 fixed (coefficient calculation function enables
Initial value memory range Same as measuring range

Measuring range, Resolution and Accuracy

Measuring range, Resolution and Accuracy Resolution Accuracy
0 to ±50000μm/m 1μm/m ±(0.05% of reading + 1) μm/m
±50000 to 500000μm/m 10 μm/m ±(0.05% of reading + 10)μm/m

Strain Measurement (High-Resolution Mode)

Constant voltage excitation Approx. 5 VDC

Constant current excitation

 

Approx. 16.7 mA (bridge resistance 350Ω)
(up to 2 km with a 4-conductor (0.5 mm 2 ) shielded cable)
Scanning speed 280 ms/channel
Gage factor 2.00 fixed (coefficient calculation function enables correction with 2.00/Ks)
Initial value memory range Same as measuring range

Same as measuring range

Measuring Range Resolution Accuracy
0 to ±20000μm/m 0.1μm/m ±(0.05% of reading + 0.3) μm/m
±20000 to 200000μm/m 1μm/m ±(0.05% of reading + 3) μm/m

 

Notes : 1. Available only with 4-gage method (bridge resistance 120 to 1000Ω)

2. Bridge resistance should be 350Ω for bridge excitation with constant current.

3. Measuring range is 0 to 15000 μm/m for bridge excitation with constant current.

4. Available only with dedicated scanners.

 

Strain Measurement(High-Speed Mode)

Bridge excitation

 

 

 

 

Constant voltage excitation: Approx. 2 VDC
Constant current excitation:
Approx. 5.7 mA (bridge resistance 350Ω )
(up to 5 km with a 4-conductor (0.5 mm 2 ) shielded cable)
Approx. 16.7 mA (bridge resistance 120Ω )
(up to 2 km with a 4-conductor (0.5 mm 2 ) shielded cable)
Scanning speed 20 ms/channel
Gage factor 2.00 fixed (coefficient calculation function enables correction with 2.00/Ks)
Initial value memory range Same as measuring range

Measuring range, Resolution and Accuracy

Measuring Range Resolution Accuracy
0 to ±50000 μm/m 1 μm/m ±(0.08% of reading + 3) μm/m
±50000 to 500000μm/m 10μm/m ±(0.08% of reading + 30) μm/m

 

Notes : 1. Available only with 4-gage method

2. Available only with dedicated scanners.

Voltage Measurement (Stondard Mode)
Scanning speed 50 ms/channel
Initial value memory range Same as measuring range

 

Measuring range, Resolution and Accuracy

Range Mode Measuring Range Resolution Accuracy Input Resistance
V/500mV 0 to ±50.000 mV 1μV ±(0.05% of reading + 3)

 

10 M Ω

or more

 

±50.00 to 500.00 mV 10μV
V/50V 0 to ±5.0000 V 100μV ±(0.05% of reading + 2)

 

1 M Ω

or more

 

±5.000 to 50.000 mV 1mV
Voltage Measurement (High-Speed Mode)
Scanning speed 20 ms/channel
Initial value memory range Same as measuring range

Measuring range, Resolution and Accuracy

Range Mode Measuring Range Resolution Accuracy Input Resistance
V/500mV ±(0.08% of reading + 6) 1μV ±(0.08% of reading + 6)

 

10 M Ω

or more

 

±50.00 to 500.00 mV 10μV
V/50V 0 to ±5.0000 V 100μV

 

1 M Ω

or more

 

±5.000 to 50.000 mV 1mV
Current Measurement (Standard Mode)
Scanning speed 50 ms/channel
Initial value memory range Same as measuring range

Measuring range, Resolution and Accuracy

Channel Mode Measuring Range Resolution Accuracy
I/50mA 0 to ±50.00mA 10μA (±0.05% of reading + 0.01) mA

 

Notes : 1. External shunt resistor (high-accuracy 250Ω ) is required.

2. Stated accuracy does not include the external shunt resistor.

 

Current Measurement (High-Speed Mode)
Scanning speed 20 ms/channel
Initial value memory range Same as measuring range

Measuring range, Resolution and Accuracy

 

Channel

Mode

 

Measuring Range Resolution Accuracy
I/50mA 0 to ±50.00 mA 10μA ±(0.08% of reading + 0.01) mA

 

Notes : 1. Available only with dedicated scanners.

2. External shunt resistor (high-accuracy 250Ω) is required.

3. Stated accuracy does not include the external shunt resistor accuracy.

 

Temperature Measurement with Thermocouples (Standard Mode)
Scanning speed 50 ms/channel

Measuring range,Resolution and Accuracy

Type Measuring Range Resolution Accuracy Internal Reference Junction Compensator Accuracy
K -200.0 to 1230.0 ℃ 0.1℃ ±0.7℃

 

±0.5 ℃

(with input terminal temperature balanced in an ambient )(temperature range of 0 to 50 ℃)

 

T -200.0 to 400.0 ℃ ±0.7℃
E -200.0 to 660.0 ℃ ±0.5℃
J -200.0 to 870.0 ℃ ±0.6℃
R -0 to 1760.0 ℃ ±2.2℃

 

Notes : 1. Accuracies do not include the internal reference junction

compensator accuracy.

2. The reference junction compensator is switchable between internal

and external.

3. Thermocouple resistance should be 1 kΩor less.

Temperature Measurement with Civil Engineering Transducers with Temperature Measuring Function (Standard Mode)
Scanning speed 50 ms/channel

 

Measuring range, Resolution and Accuracy

Measuring Range Resolution Accuracy
-50.0 to 200.0℃ 0.1℃ ±0.5℃

 

Notes : 1. Target physical quantity and temperature can be measured

in a single channel.

2. Strain measuring ranges are the same as in strain measurement

in standard mode.

 

Temperature Measurement with Platinum Resistance Thermometer Bulb (Standard Mode)
Scanning speed 50 ms/channel

Measuring range, Resolution and Accuracy

Type Measuring Range Resolution Accuracy
Pt100 -200.0 to 660.0℃ 0.1℃ ±0.3℃
JPt100 -200.0 to 510.0℃

Note : Connection is 3-wire system

Measurement with Potentiometer Sensor
Scanning speed

50 ms/channel (standard mode)

20 ms/channel (high-speed mode)

Initial value memory range Same as measuring range
Sensor power supply  Approx. 2 VDC
Potentiometer resistance 1 to 10 kΩ

Measuring range, Resolution and Accuracy

Channel Mode Measuring Range Resolution Accuracy
POT. 0 to ±50.00% 0.01% ±0.1%FS

Note : Connection is 3-wire system

Clock (UCAM-60B)

Real-time clock is built in

(battery backup 5 years).

Display (UCAM-60B) Fluorescent display tube,
128 x 64 dots Printer (UCAM-60B)
Printing system Thermal
Paper width : 58 mm (24 characters/line), UCAM-60A-RP
Printing speed : 60 mm/sec
PC Card Slot Conforms with PCMCIA Ver. 4.2.
Accepts a commercially available flash ATA card or
CF card (ATA card adapter required).
Interfaces RS-232C and LAN (10BASE-T/100BASE-TX)
File Conversion CSV conversion
Self-diagnosis Function

Checks display, printer, bridge excitation, leadwire-off, input/output

resistance, insulation resistance, mode, etc. Checking of input/outut

resistance and mode is availabe only for dedicated scanners.

TEDS Compatibility Interface : IEEE 1451.4 Mixed Mode Transducer Interface Class 2
Applicable sensor : Should have information written in accord-ance with IEEE template No. 33; cable length should be 30 m or less.
Operating Temperature & Humidity Range: 0 to 50℃, 20 to 85% RH (noncondensing)
Power Supply AC85 to 264V, 50/60Hz (AC-operated version)
10 to 16 VDC (DC-operated version)
Note : DC-operated version has power control function.
Current Consumption 0.5 A or less: 100 VAC (with 3 dedicated scanners mounted)
4 A or less: 12 VDC (with 3 dedicated scanners mounted)
Demensions UCAM-60B: 360(W) x 88(H) x 400(D) mm (excluding protrusions)
UCAM-65B: 327(W) x 88(H) x 365(D) mm (excluding protrusions)
Weight UCAM-60B: Approx. 8 kg,UCAM-65B: Approx. 4.6 kg

 

Standard Accessories

 

AC power cable P-18 with conversion adapter CM-39 (AC-operated version) DC power cable P-57 (DC-operated version)

Recording paper UCAM-60A-RP (1 roll for UCAM-60B only)

Screwdriver, Spare fuse, Instruction Manual

CD-R (Control Software UCS-60B for UCAM-65B only)

 

Optional Accessories

Recording Paper UCAM-60A-RP (10 rolls/pack)

Dedicated Scanners

USS-61B/62B/63B
Models USS-61B (TEDS compatible)
USS-62B (with NDIS connectors, TEDS compatible)

 

USS-63B (for civil engineering measurement,TEDS compatible, with lightning arrester)

 

Number of Measuring Channels 10/unit
Switching Terminals Semiconductor relays
Input Terminals Can connect to leadwires through either soldering or screwing.
NDIS connectors (USS-62B)
One-touch terminal block JT-1A (optional)
Lightning Arrester Provided (USS-63B only)
Operating Temperature & Humidity Range 0 to 50℃, 20 to 85% RH (noncondensing)
Dimensions 320(W) x 28(H) x 80(D) mm (excluding protrusions)
Weight USS-61B : Approx. 800 g (including terminal cover)
USS-62B : Approx. 1 kg (including terminal cover)
USS-62B : Approx. 1 kg (including terminal cover)

Standard Accessories

Terminal cover, Channel label and for USS-62B, NDIS connector caps (pre-attached to connectors)

Scanner Interfaces

USI-67A
Connectable Scanners USB-70 series
Number of Connectable Scanners Max. 20
Operating Temperature & Humidity Range 0 to 50℃, 20 to 85% RH (noncondensing)
Dimensions 99(W) x 50(H) x 163(D) mm (excluding protrusions),
Weight Approx. 160 g

USI-67A

 

External Input/Output Unit

UIO-60A

Output:

 

ALARM signal 4 channels (high/low limit checking)
BUSY signal 1 channel

Input:

 

 

 

 

 

 

START signal 1 channel
STOP signal 1 channel
RESET signal 1 channel
RAINFALL signal 1 channel
Operating Temperature & Humidity Range 0 to 50℃, 20 to 85% RH (noncondensing)
Dimensions 90(W) x 50(H) x 180(D) mm(excluding protrusions),
Weight Approx. 140g

External Scanners

USB-70B
Models USB-70B-10 (for general strain measurement)
  USB-70B-20 (for general strain meas., with NDIS connectors)
  USB-70B-30 (for civil engineering, with lightning arrester)
Number of Measuring Channels 50/unit
Measuring Channel Mode Selected for each channel from the mainframe
Connectable Sensors USB-70B-10 Strain gages, strain gage transducers,potentiometer Sensors,DC voltage-output instruments, thermocouples
USB-70B-20 Strain gages, strain gage transducers,potentiometer Sensors,DC voltage-output instruments, thermocouples(trans-ducers with NDIS connector can be connected)
USB-70B-30 Strain gages, strain gage transducers,potentiometer Sensors,DC voltage-output instruments,thermal Sensors (thermo-couples, platinum resistancethermometer bulbs, civil engineering transducers withtemperature measuring function);lightning arrester built in
Power Supply Supplied from data logger If the cable is extended
or if 4 or more scanners are connected, an optional
UPS-70B should be mounted into scanners.
UPS-70B operates on 90 to 132/180 to 264 VAC
(no switchover required), 50/60 Hz
Operating Temperature & Humidity Range 0 to 50℃, 20 to 85% RH (noncondensing)
Dimensions 302(W) x 107(H) x 500(D) mm (excluding protrusions)
Weight Approx. 7.7 kg (USB-70B-30)
Approx. 8.5 kg (USB-70B-20)

 

Quarter bridge method